factory set, located on the Main Board. According to the instrument power supply
configuration, the breaker ON/OFF main switch changes as follows:
The high voltage section is provided of a series of protection fuses as reported in
the following table:
Power Supply
Breaker
120 Vac
15A
230 Vac
7A
Fuse
Type
Protection
F1
F2A; IEC 127/I; (5 x20 mm)
Injector heater element
F2
T1.6A; IEC 127/III; (5 x20 mm)
230 Vac Main power (Breaker)
120 Vac Main power (Breaker)
F3
F2A; IEC 127/I; (5 x20 mm)
Auxiliary heater element
F4
F0.63A; IEC 127/I; (5 x 20 mm)
Oven fan motor
F5
F0.63A; IEC 127/I; (5 x 20 mm)
Oven fan motor
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...