Chapter 8
Detector Signal Parameters
Flame Ionization Detector (FID)
169
OPERATING SEQUENCE
Setting the FID Signal Parameters
1. In Main Menu move the cursor until
Signal
is selected. Press
ENTER
to
display the
SIGNAL
menu.
2. Scroll to
Range 10^ (0...3)
and set the electrometer amplifier input
range. The value 0 (10
0
) is the most sensitive.
3. Scroll to
Run
Autozero
and set On.
4. If offset is required, scroll to
Offset
and enter a numeric value or set it On
to recall the last offset from memory.
NOTE
If the Range 10^ is set 2 or 3, the small variation of the output signal is not detected. For
this reason, the, Signal pA, Ign. thresh and Flameout retry parameters will be not
displayed in the
DETECTOR
menu.
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...