Chapter 14
SOP number: P0474/05/E - 02 May 2007
256
Table 14-3. TCD-PPKD Acceptance Criteria
Accep
tance V
alues
CHROM-CARD
Baseline Parameters
Analog (1V Full Scale)
Digital (10V Full Scale)
Noise (
μ
V)
< 30
< 300
Wander (
μ
V)
< 140
< 1 400
Drift (
μ
V/h)
< 200
< 2 000
Analytical Results
Analog (1V Full Scale)
Area Counts (0.1
μ
Vs)
Digital (10V Full Scale)
Area Counts (0.1
μ
Vs)
Components
> 30 000
for each component
> 300 000
for each component
Using Chrom-Card, connected to the FOCUS GC digital output, refer to Analog or Digital Acceptance Values
according to the TRACE Signal Time (0 or 1) set in Chrom-Card WCC.INI Configuration.
Computing-integrator (e.g. ChromJet)
When a Computing-integrator (e.g. ChromJet) is used, the peak area counts will result to be 5 times lower than the
peak area counts obtained by using Analog Chrom-Card.
Accep
tance V
alues
CHROMQUEST
Baseline Parameters (1V Full Scale)
Noise (
μ
V)
< 30
Wander (
μ
V)
< 140
Drift (
μ
V/h)
< 200
Analytical Results (1V Full Scale) - Area Counts (0.01
μ
Vs)
Components
> 300 000 for each component
Using Chrom-Quest, connected to the FOCUS GC digital output, and using a range of zero, set in the ChromQuest
configuration page, a multiplier of 10
-8
in
V scale (or 10
-5
in mV scale) in order to have equivalency with FOCUS
GC displayed signal output. Different range values need a change of the multiplier used in ChromQuest.
Analytical Acceptance Comments
1
In case of TCD equipped with the polyimide coated filaments, the acceptance values of the Components Area will
result to be 4 times lower than the values reported in Table 14-3.
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...