Chapter 15
Maintenance and Troubleshooting
Maintaining a PPKD Injector
280
Removing the Liner
4. Unscrew the nut that retains the analytical column, then remove the column
with the inserted ferrule. See A of Figure 15-10
5. Unscrew the retaining nut at the bottom of the injector. Remove the adapter
and the liner then remove the ferrule. See B of Figure 15-10.
Figure 15-10. PPKD Injector Replace/Clean the Liner
6. If you are using a new liner, go to step 8. If you are cleaning the liner, put the
dirty liner into an ultrasonic bath filled with a methanol/acetone mixture (1:1)
and clean it for about half an hour.
7. Using tweezers, remove the liner from the bath and dry it with compressed
clean air.
PRECAUTIONS
A B C D
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...