Chapter 14
SOP number: P0474/05/E - 02 May 2007
254
19. In
SIGNAL
menu
select Run auto zero
, then press
ENTER
to perform
the automatic zeroing of the system.
Baseline acquisition and Analysis
NOTE
Refer to the Acceptance Values reported in theTable 14-3 according to the data handling in
use.
20. Perform a blank analysis without injecting and press
START
on the GC to
begin the checkout run.
21. With the GC in Stand-by/Prep Run condition, activate the data system for 10
minutes to evaluate your baseline in isothermal condition.
22. After the baseline evaluation has been completed, set-up the data system to
acquire a single run.
23. Inject the test mixture and press
START
on the GC to begin the checkout run.
The resulting chromatogram should look like the one shown in Figure 14-1.
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...