Chapter 8
Flame Ionization Detector (FID)
Detector Signal Parameters
168
6. Refer to
Setting the FID Signal Parameters
instructions on setting the signal parameters.
If you desired to check or modify the detector gas flows operate as follows:
•
Set the detector base body temperature
Off
.
•
Loosen the fixing screw on the front of the detector cell and remove the
detector.
•
Turn the air and make-up gas flows
Off
.
•
Turn the
Hydrogen
flow
O
n.
•
Connect the flow meter to the detector base body using the adapter.
•
Measure the gas flow and verify that the flow rate is 35 mL/min.
•
By using a screwdriver, adjust the gas flow with the flow regulator until the
desired gas flow is achieved.
•
Turn the hydrogen flow
Off
.
WARNING! Never measure air and hydrogen flow together.
•
Turn the
Air
supply
O
n.
•
Measure the gas flow and verify that the flow rate is 350 mL/min.
•
By using a screwdriver, adjust the gas pressure with the pressure regulator
until the desired gas flow is achieved.
•
Turn the
Air
flow
Off
.
•
Turn the
Makeup
gas flow
O
n.
•
Measure the gas flow and verify that the flow rate is 30 mL/min.
•
By using a screwdriver, adjust the gas flow with the pressure flow until the
desired gas flow is achieved.
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...