Chapter 3
Detector Gas Regulation
Gas Control
77
Verify that the corresponding gases flows are correct. If not adjust the value
turning the relevant regulation screw as shown in Figure 3-2.
Figure 3-2. TCD Detector Gases Flow Regulation
The flow is measured at the exit of the detector. Enter the TCD detector control
menu to turn the flow on/off.
Gas
Flow
Helium (make-up gas)
27 mL/min
Helium (reference gas)
30 mL/min
1 turn clockwise of the regulation screw increases the flow of about 2.5 mL/min
1 turn counterclockwise of the regulation screw decreases the flow of about 2.5 mL/min
Ref.
Make-up
OPEN
TCD Detector
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...