Chapter 7
Column
Keeping Leaks Under Control
148
OPERATING SEQUENCE
Performing a Column Evaluation
The column flow control in the FOCUS GC is indirect. This means that the GC
regulates pressure to control the flow of gas through the column. To do this, the
GC relies on a column constant. The column constant is a measure of the columns
pneumatic resistance. Use the FOCUS GC’s column evaluation feature to
automatically calculate the column constant.
To perform the column evaluation, the system uses the correlation between the
applied pressure to the flow and the column temperatures. This operation must be
carried out every time a new column is installed. Before performing the column
evaluation, a manual checking for leaks of the system must be carried out.
The GC must not be performing a run and must be isothermally stable before you
can perform a column evaluation.
1. In Main Menu move the cursor until
Carrier
is selected, then press
ENTER
.
2. Scroll to
Column evaluation
and press
ENTER
to open
COLUMN
EVALUATION
menu.
3. To start column evaluation, scroll to
Run column eval
and press
ENTER
.
NOTE
To exit Column evaluation menu without performing column evaluation, scroll to Exit w/o
column evaluation and press ENTER.
CARRIER (He)
Column evaluation
COLUMN EVALUATION
Run column eval.
K = 1.0
Exit w/o column eval
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...