Functional Test Procedures
Functional Test Procedures
There are three procedures in this section that provide a quick way to con
fi
rm
basic functionality and proper adjustment of the DSA8300:
Instrument and Module Compensation
These procedures require little or no test equipment. For information on when to
use these procedures, refer to Performance Veri
fi
cation Procedures. (See page 1,
Instrument Diagnostics
The following steps run the internal routines that con
fi
rm basic functionality
and proper adjustment.
Equipment required
None
Prerequisites
Install all sampling modules to be diagnosed: see module user
manual for installation instructions.
Power on the instrument and wait 20 minutes (warm-up period)
before doing this procedure.
2
DSA8300 Performance Veri
fi
cation