![Tektronix 80A00 Скачать руководство пользователя страница 200](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946200.webp)
Electrical Sampling Modules
Table 3: Data for calculation of gain and linearity
V
in
mV
Input V
in
Measured
V, mV
x
x
2
y
xy
Y = ax+b
Deviation lower
limit
upper
limit
-450
-10
+10
-350
-10
+10
-250
-10
+10
-150
-10
+10
-50
-10
+10
50
-10
+10
150
-10
+10
250
-10
+10
350
-10
+10
450
-10
+10
-10
+10
n
S
x
S
xx
S
y
S
xy
10.
Change all settings in the setup to Channel 2, (except 80E06 and 80E11X1)
and then repeat steps 1 through 9 until all DC vertical voltage accuracy, single
point, compensated calculations listed in the table are checked. (See Table 3.)
11.
Disconnect the test setup.
End of Procedure
Random Noise, Displayed
These procedures check those characteristics that relate to the signal acquisition
and are listed as checked in the
DSA8300 Digital Serial Analyzer Speci
fi
cations
manual.
Equipment required
No external equipment required
Prerequisites
182
DSA8300 Performance Veri
fi
cation