![Tektronix 80A00 Скачать руководство пользователя страница 105](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946105.webp)
Performance Tests
80C10C performance test
Minimum
Incoming
Outgoing
Maximum
11602
–1.02
–1.87
–0.17
15469
–1.86
–2.71
–1.01
19336
–3.00
–3.86
–2.16
23203
–4.51
–6.19
–2.83
25781
–5.71
–7.87
–3.55
27070
–6.37
–8.75
–3.99
30938
–8.54
–11.53
–5.56
34805
–10.93
–14.45
–7.41
38672
–13.41
–17.41
–9.41
40GBase-FR (41.25 Gb/s, ENET41250) Reference Receiver setting frequency response (dB) — OPTION F1 and OPTION
F3 ONLY
Frequency (GHz)
Nominal response (dB)
0.00
0.00
–0.85
+0.85
6188
–0.11
–0.96
+0.74
12375
–0.45
–1.30
+0.40
18563
–1.02
–1.87
–0.17
24750
–1.86
–2.71
–1.01
30938
–3.00
–3.86
–2.16
37125
–4.51
–6.19
–2.83
41250
–5.71
–7.87
–3.55
43313
–6.37
–8.75
–3.99
49500
–8.54
–11.53
–5.56
55688
–10.93
–14.45
–7.41
61875
–13.41
–17.41
–9.41
Aberrations (peak to peak)
In
fi
niband EDR setting
none
5%
100GBase-R4 setting
none
5%
OTU4
none
5%
28.05 GHz
none
5%
32 GHz
none
5%
OC-768/STM-256
none
5%
40GBase-FR
none
5%
FEC43.02, OTU3
none
5%
55 GHz
none
10%
70 GHz
none
12%
Filter or bandwidth
80 GHz
none
12%
DSA8300 Performance Veri
fi
cation
87
Содержание 80A00
Страница 2: ......
Страница 6: ......
Страница 16: ...Important safety information x DSA8300 Performance Verification ...
Страница 213: ...Electrical Sampling Modules Figure 21 Setup to measure Power in DSA8300 Performance Verification 195 ...