![Tektronix 80A00 Скачать руководство пользователя страница 175](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946175.webp)
DSA8300 Performance Veri
fi
cation
Setup
Figure 8: External direct trigger sensitivity test setup
1.
Install the 80E0X module(s) to test starting at the left-most small module
slot as shown in the
fi
gure. Power on the instrument and wait 20 minutes
for warm-up.
2.
Connect the test equipment (See Figure 8.)
3.
Set the synthesized signal generator frequency as follows:
Frequency to
2.5 GHz
(400 ps period).
Output to
400 mV
p-p
(200 mV displayed).
4.
Set the DSA8300 as follows:
a.
Push the
Default Setup
front-panel button and click
Yes
.
b.
Select
C1
from the Waveform Selector menu (left end of controls bar
at bottom of the graticule).
c.
Select
Setup > Mode/Trigger
and select
Direct
for Trigger Source.
d.
Click the
Vert
tab and set the following:
Scale to
20 mV/div
.
Offset to
0
.
e.
Click the
Horz
tab and set the following:
Scale to
200 ps/division
.
Position to
19 ns
(29 ns if using 80E07B, 80E08B, 80E09B, 80E10B).
Horizontal Reference to
0%
.
DSA8300 Performance Veri
fi
cation
157