![Tektronix 80A00 Скачать руководство пользователя страница 202](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946202.webp)
Electrical Sampling Modules
8.
Record the noise (V STD DEV) reading.
9.
Check noise V for the module tested in the following table.
Table 4: Random noise limits
Module
Bandwidth setting
Noise limit
80E01
≤
2.3 mV
80E02
≤
800
μ
V
80E03, 80E04
≤
1.2 mV
80E06
≤
2.4 mV
30 GHz BW setting
≤
410
μ
V
80E07B, 80E08B
20 GHz BW setting
≤
380
μ
V
60 GHz BW setting
≤
600
μ
V
40 GHz BW setting
≤
480
μ
V
80E09B
30 GHz BW setting
≤
410
μ
V
50 GHz BW setting
≤
700
μ
V
40 GHz BW setting
≤
480
μ
V
80E10B
30 GHz BW setting
≤
410
μ
V
70 GHz
<1100
μ
V
60 GHz
<600
μ
V
80E11, 80E11X1
30 GHz
>480
μ
V
10.
Repeat steps 1 through 8 for each bandwidth setting.
11.
Repeat steps 1 through 8 for channel 2.
End of Procedure
Analog Bandwidth up to 50 GHz (80E01, 80E06, 80E07B, 80E08B, 80E09B,
80E10B, 80E11, 80E11X1)
These procedures check those characteristics that relate to the signal acquisition
and are listed as checked in the
DSA8300 Digital Serial Analyzer Speci
fi
cations
manual. This check has two procedures that characterizes the reference signals.
The last procedure characterizes the sampling module.
NOTE.
This test setup is limited to 50 GHz. To test modules with a bandwidth
above this, you will also need to perform the Analog Bandwidth to 60 GHz
procedure.