![Tektronix 80A00 Скачать руководство пользователя страница 179](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946179.webp)
DSA8300 Performance Veri
fi
cation
20.
Click the
Acq
tab and set Stop After to
Run/Stop Button Only
.
21.
Push
RUN/STOP
until it toggles to green. Position the rising edge of the
waveform so that it goes through the center of the display graticule.
22.
Set Stop After to
Condition
(Acq tab).
23.
Push
CLEAR DATA
, push
RUN/STOP
, and wait for the acquisition to stop.
24.
Verify that the RMS Jitter measurement readout is less than
2.5 ps
.
25.
Click the
Horz
tab and set the Timebase Position parameter to
1
μ
s
26.
Click the
Acq
tab and set Stop After to
Run/Stop Button Only
.
27.
Push
RUN/STOP
until the button color toggles to green. Position the rising
edge of the waveform so that it goes through the center of the display
gratucule.
28.
Set Stop After to
Condition
(in the Acq tab).
29.
Push
CLEAR DATA
, push
RUN/STOP
, and wait for the instrument to stop
acquisition.
30.
Verify that the RMS Jitter measurement is less than
11.5 ps
.
31.
Move the signal source cable to the next module under test. Repeat from
step for all 80E00 series modules installed. Check one input per module (for
example, channels 1 and 3 if two modules are installed).
32.
In Acq Setup dialog box, set Stop After to
Run/Stop Button Only
when
fi
nished.
End of Procedure
Trigger Delay Jitter, Untriggered Phase Corrected Mode
Equipment required
82A04 Phase Reference module
0.5 meter precision SMA cable (item 15)
1.0 meter precision SMA cable (item 16)
Power divider (item 26)
Synthesized signal generator (item 31)
Electrical sampling module (80E01, 80E02, 80E03, 80E04,
80E06, 80E07B, 80E08B, 80E09B, or 80E10B)
Prerequisites
As listed under
Prerequisites
DSA8300 Performance Veri
fi
cation
161
Содержание 80A00
Страница 2: ......
Страница 6: ......
Страница 16: ...Important safety information x DSA8300 Performance Verification ...
Страница 213: ...Electrical Sampling Modules Figure 21 Setup to measure Power in DSA8300 Performance Verification 195 ...