![Tektronix 80A00 Скачать руководство пользователя страница 218](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946218.webp)
Electrical Sampling Modules
c.
Read the power loss from the SNA screen, and record it for each frequency
in the
Adapter Loss
column of the table. (See Table 9.)
d.
Turn Trace Hold on the SNA to
Off
.
3.
Disconnect the Waveguide adapters (separating them from each other), the
Waveguide power sensor, and the Power meter from the test setup. Leave the
remainder of the test setup intact for use in the next procedure.
HF Setup and Test
Do the following procedure to set up and test the HF Response.
NOTE.
Use a torque wrench when connecting the terminators, adapters,
attenuator, and Waveguide power sensor in this setup. Use 8 ±0.3 inch pounds
of torque.
1.
Refer to the following
fi
gure as you perform the following substeps:
a.
Start with the remainder of the test setup from step 3 in the last procedure.
b.
If testing an 80E06, install the 80N01 or 80X01 Extender Cable in the
leftmost slot of the main instrument and connect the cable end of the
Extender to the 80E06. If testing an 80E09B, install the module in the
leftmost slot.
c.
Connect the one end of a Waveguide-to-1.85 mm male adapter to the
input channel of the module to be tested. Connect the other end to the
Waveguide section.
200
DSA8300 Performance Veri
fi
cation
Содержание 80A00
Страница 2: ......
Страница 6: ......
Страница 16: ...Important safety information x DSA8300 Performance Verification ...
Страница 213: ...Electrical Sampling Modules Figure 21 Setup to measure Power in DSA8300 Performance Verification 195 ...