![Tektronix 80A00 Скачать руководство пользователя страница 226](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946226.webp)
Electrical Sampling Modules
f.
In the Meas Setup dialog box:
Select
Meas 1
and make sure the
Source1
button is selected.
Make sure the channel under test is selected in the Source tab, click
the
Pulse
Signal Type button, and then choose
Select Meas > Pulse –
Amplitude > Max
.
Select the
Region
tab.
Set Gate
G2
to
46%
.
Turn Gates
On
(check).
g.
In the Meas Setup dialog box:
Select
Meas 2
and make sure the
Source1
button is selected.
Make sure the channel under test is selected in the Source tab, click
the
Pulse
Signal Type button, and then choose
Select Meas > Pulse –
Amplitude > Min
.
Select the
Region
tab.
Set Gate
G2
to
46%
.
Set Gates
On
(check).
h.
The Max and Min measurements are now set to include the region of
250 ps to 20 ps before the edge. The aberration speci
fi
cation over this
region is ±3%, which is ±30 m
ρ
for a 1
ρ
step. Since the reference is –1
ρ
,
the Max and Min values should be within –1.030 to –0.970
ρ
to meet
the ±3% speci
fi
cation.
i.
Calculate and take down for later use the peak aberrations in percent using
the formulas (with Max and Min in m
ρ
units, not
ρ
):
Positive aberration (%) = (1000+Max) * 0.1
Negative aberration (%) = (1000+Min) * 0.1
For example: if Max = –989 m
ρ
and Min = –1.003
ρ
,
positive aberration (%) = (1000 + (–989)) * 0.1 = +1.1
negative aberration (%) = (1000 + (–1003)) * 0.1 = –0.3
NOTE.
The aberration can have either a positive or negative sign. Both
aberrations could have the same sign when the waveform does not cross
the reference line in the region being checked.
208
DSA8300 Performance Veri
fi
cation