![Tektronix 80A00 Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946027.webp)
Functional Test Procedures
For the 80E01, 80E02, 80E03, 80E04, and 80E06
Limits
Vertical Scale
(mV/div)
Vertical Offset
(mV)
DC CAL Output
(mV)
Minimum (V)
Maximum (V)
100
–1000.0
–1000.0
–1.0090
–.9910
100
0.0
–450
–461.0
–439.0
100
0.0
0
–2.0
2.0
100
0.0
450
439.0
461.0
100
1000.0
1000.0
991.0
1009.0
For the 80E07B, 80E08B, 80E09B, and 80E10B
Limits
Vertical Scale
(mV/div)
Vertical Offset
(mV)
DC CAL Output
(mV)
Minimum (mV)
Maximum (mV)
100
–600.0
–600.0
–605.4
–594.6
100
0.0
–450
–461.0
–439.0
100
0.0
0
–2.0
2.0
100
0.0
450
439.0
461.0
100
600.0
600.0
594.6
605.4
10.
Test all channels:
Repeat steps 3 through 9 until all input channels are veri
fi
ed.
11.
Remove the test setup:
Disconnect the SMA cable from the channel input
and the DC CALIBRATION output.
DSA8300 Performance Veri
fi
cation
9