Electrical Sampling Modules
d.
Record the reading on the power meter in the
Reference Level
column.
e.
Turn Trace Hold on the SNA to
Off
.
5.
Leave the test equipment hooked up for use in the steps under
Adapter
Characterization
.
Adapter Characterization
Do the following procedure to derive the power loss of the adapters:
1.
Refer to the following
fi
gure as you perform the following substeps:
a.
Start with the test setup as left from step 5 in the last procedure.
b.
Disconnect the Waveguide power sensor from the Waveguide section.
c.
Connect the 1.85 mm end of a Waveguide-to-1.85 mm female adapter to
the 1.85 mm end of a Waveguide-to-1, 85 mm male adapter.
d.
Install the set of combined adapters between the Waveguide power sensor
and the Waveguide section as is shown in the following
fi
gure.
198
DSA8300 Performance Veri
fi
cation