Functional Test Procedures
1.
Initialize the instrument:
Push the front-panel
DEFAULT SETUP
button and
click
Yes
in the con
fi
rmation dialog box.
2.
Set the Trigger System:
In the UI application toolbar, select
Free Run
from
the Trig menu.
3.
Connect the SMA cable from the DC CALIBRATION output to the channel
input to test.
4.
Set the DC CALIBRATOR OUTPUT:
a.
Push the front-panel Vertical
MENU
button.
NOTE.
When an optical module is installed, the optical setup dialog box
displays by default. Click the
Basic
button to display the basic dialog box.
b.
Set the DC CAL value to
200 mV
.
c.
Push the front-panel Vertical
MENU
button again to dismiss the Vert
Setup dialog box.
5.
Push the channel button for the channel to test. The button lights and the
screen shows the selected channel signal.
Figure 2: Channel button location
DSA8300 Performance Veri
fi
cation
7