![Tektronix 80A00 Скачать руководство пользователя страница 274](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946274.webp)
80A05 Electrical Clock Recovery Module
h.
Make sure the Meas 1
ON
check box is selected.
i.
Verify that
Use Wfm Database
is selected.
Test
Follow this procedure to make the clock recovery sensitivity and clock timing
jitter checks:
1.
In the Mode/Trigger dialog box, select the clock recovery rate for C1 that you
want to test as shown in the following table.
2.
Set the D3186 Pulse Pattern Generator to the rate setting and amplitude for the
clock recovery rate that you selected in step 1 as shown in the following table.
3.
Adjust the Horizontal SCALE knob on the front panel of the instrument to
the setting listed in the table for the clock recovery rate that you selected in
step 1 as shown in the following table.
Table 14: Clock recovery settings
D3186
Clock recovery rate
Rate setting
Amplitude
Horizontal Scale setting
OC-3
0.15552 Gb/s
1.0 V
1 ns/div
OC-12
0.62208 Gb/s
1.0 V
500 ps/div
FC-1063
1.0625 Gb/s
1.0 V
200 ps/div
GBE
1.25 Gb/s
1.0 V
200 ps/div
Serial ATA
1.50 Gb/s
1.0 V
200 ps/div
FC2125
2.125 Gb/s
1.0 V
100 ps/div
OC-48
2.48832 Gb/s
1.0 V
100 ps/div
2GBE
2.5 Gb/s
1.0 V
100 ps/div
In
fi
niband
2.5 Gb/s
1.0 V
100 ps/div
PCI Express
2.5 Gb/s
1.0 V
100 ps/div
2.5 G G.709 FEC
2.666 Gb/s
1.0 V
100 ps/div
Serial ATA
3.0 Gb/s
1.5 V
100 ps/div
XAUI
3.125 Gb/s
1.5 V
100 ps/div
FibreChannel
3.188 Gb/s
1.5 V
100 ps/div
VSR5
3.318 Gb/s
1.5 V
100 ps/div
4 Gigabit FibreChannel
4.25 Gb/s
1.5 V
100 ps/div
Serial ATA
1
6.0 Gb/s
1.5 V
200 ps/div
XAUI
1
6.25 Gb/s
1.5 V
100 ps/div
OC-192
1
9.95328 Gb/s
1.5 V
50 ps/div
10GBASE-W
1
9.95328 Gb/s
1.5 V
50 ps/div
10GBASE-R
1
10.3125 Gb/s
1.5 V
50 ps/div
10GFC
1
10.51875 Gb/s
1.5 V
50 ps/div
G.975 FEC
1
10.66 Gb/s
1.5 V
50 ps/div
256
DSA8300 Performance Veri
fi
cation
Содержание 80A00
Страница 2: ......
Страница 6: ......
Страница 16: ...Important safety information x DSA8300 Performance Verification ...
Страница 213: ...Electrical Sampling Modules Figure 21 Setup to measure Power in DSA8300 Performance Verification 195 ...