![Tektronix 80A00 Скачать руководство пользователя страница 116](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946116.webp)
Performance Tests
80C11B performance test
Minimum
Incoming
Outgoing
Maximum
11245.0
–6.37
–8.75
–3.99
12851.1
–8.54
–11.53
–5.56
14457.7
–10.93
–14.45
–7.41
16064.4
–13.41
–17.41
–9.41
OC-192/STM-64 (9.953 Gb/s) Reference Receiver setting frequency response (dB)
Frequency (MHz)
Nominal response (dB)
0.00
0.00
–0.85
+0.85
1493.2
–0.11
–0.96
+0.74
2986.0
–0.45
–1.30
+0.40
4478.8
–1.02
–1.87
–0.17
5972.4
–1.86
–2.71
–1.01
7465.0
–3.00
–3.86
–2.16
8958.0
–4.51
–6.19
–2.83
9953.28
–5.71
–7.87
–3.55
10451.2
–6.37
–8.75
–3.99
11944.0
–8.54
–11.53
–5.56
13437.2
–10.93
–14.45
–7.41
14930.4
–13.41
–17.41
–9.41
Clock Recovery Optical Sensitivity — 80C11B-CR1 ONLY
Minimum Average Optical Input Power
–7.5 dBm
none
Clock Recovery Optical Sensitivity — 80C11B-CR2 ONLY
Minimum Average Optical Input Power
–7.5 dBm
none
Clock Recovery Optical Sensitivity — 80C11B-CR3 ONLY
Minimum Average Optical Input Power
–7.5 dBm
none
Clock Recovery Optical Sensitivity — 80C11B-CR4 ONLY
Data rate
Sensitivity
9.8 to 11.25 G/bs
–9 dBm
ER>8.2 dB
none
Minimum Average
Optical Input Power
11.25 to 12.6 Gb/s
–7.5 dBm
ER>8.2 dB
none
Recovered clock timing jitter (80C11B-CR1 only)
Rate
OC-192
none
2 ps
RMS
Recovered clock timing jitter (80C11B-CR2 only)
Rate
OC-192
none
2 ps
RMS
10GBE FEC
none
2 ps
RMS
98
DSA8300 Performance Veri