Figure 11: Phase corrected, Clock Eye external triggered jitter test setup
1.
Install the 82A04 Phase Reference module into the right-most small module
slot of the instrument.
2.
Install the 80E0X module(s) to test into the left-most small module slots.
Power on the instrument and wait 20 minutes for warm-up.
3.
Connect the test equipment (See Figure 11.)
4.
Set the synthesized signal generator as follows:
Frequency to
10.0 GHz
.
Output to
1.6 V
p-p
(800 mV
p-p
displayed).
5.
Set the DSA8300 as follows:
a.
Push the
Default Setup
front-panel button and click
Yes
.
b.
Select the input channel of the module under test from the Waveform
Selector menu (left end of controls bar at bottom of the graticule).
c.
Click the
Mode/Trigger
tab and set the following:
Trigger Source to
Clock
and clock type to
Clock/Prescale
.
Scope Mode (Clock Trigger Source) to
Eye
.
d.
Select
Setup > Phase Ref
and set the following:
Phase correction mode to
Triggered
.
Source to
C7_C8
.
Input freq to
10.00 GHz
.