Electrical Sampling Modules
Analog Bandwidth above 50 GHz (80E06, 80E09B, 80E11, 80E11X1)
These procedures check those characteristics that relate to the signal acquisition
and are listed as checked in the
DSA8300 Digital Serial Analyzer Speci
fi
cations
manual.
This procedure checks the upper bandwidth of the 80E06 and 80E09B modules.
NOTE.
This procedure checks tests the >50 GHz to 65 GHz frequencies. The
procedure Analog Bandwidth to 50 GHz checks the lower bandwidths. (See
page 184, Analog Bandwidth up to 50 GHz (80E01, 80E06, 80E07B, 80E08B,
80E09B, 80E10B, 80E11, 80E11X1).)
Equipment required
BNC to banana adapter (item 4)
Cable, special, power/phase stability (item 14)
Cable, coaxial, SMA (item 15)
Cable, BNC (item 17)
Scalar network analyzer (SNA) (item 42)
Re
fl
ectometer, with OML DC power cable to dual banana plug
(item 44)
Power meter (item 45)
Waveguide power sensor (item 46)
Adapter, Wave Guide-to-1.85 mm male (item 51)
Adapter, Wave Guide-to-1.85 mm female (item 52)
Waveguide section (item 53)
80N01 or 80X02 Module Extender (item 55)
Prerequisites
You must have performed the procedure
Analog Bandwidth to
50 GHz
Analog Bandwidth up to 50 GHz (80E01,
80E06, 80E07B, 80E08B, 80E09B, 80E10B, 80E11, 80E11X1)
SNA & Re
fl
ectometer
Characterization
Do the following steps to store re
fl
ected reference levels in the SNA:
NOTE.
Use a torque wrench when connecting the terminators, adapters,
attenuator, and power sensor in this setup. Use 8 ±0.3 inch pounds of torque.
194
DSA8300 Performance Veri
fi
cation