![Tektronix 80A00 Скачать руководство пользователя страница 169](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946169.webp)
DSA8300 Performance Veri
fi
cation
DSA8300 Performance Veri
fi
cation
These procedures check the main instrument speci
fi
cations that are listed as
checked in the
DSA8300 Digital Serial Analyzer Speci
fi
cations
manual.
Prerequisites
Be sure you have performed the
Prerequisites
before doing any procedures in this
section. (See page 16,
NOTE.
Use a torque wrench when connecting the terminators, adapters,
attenuator, and power sensor in this setup. Use 8 ±0.3 inch pounds of torque.
NOTE.
Throughout these procedures, when steps call for the display of setup
dialog boxes, you can display them from the Setup menu in the menu bar at the
top of the screen. These procedures assume you will display and dismiss the
setup dialog boxes as needed.
NOTE.
The procedures throughout this chapter contain instructions based on the
menus and controls from the most updated instrument software available at the
time this document was released. These procedures will work for other versions of
software, but some control and menu names may vary slightly.
Time Interval Accuracy, Direct and Clock Trigger
Equipment required
0.5 meter precision SMA cable (item 15)
1.0 meter precision SMA cable (item 16)
Power divider (item 26)
Synthesized signal generator (item 31)
Electrical sampling module (80E01, 80E02, 80E03, 80E04,
80E06, 80E07B, 80E08B, 80E09B, or 80E10B)
Prerequisites
As listed under
Prerequisites
DSA8300 Performance Veri
fi
cation
151