Electrical Sampling Modules
Figure 24: HF response setup
2.
Set up the main instrument as follows:
a.
Push the
DEFAULT SETUP
front-panel button and click
Yes
.
b.
Select
C1
from the Waveform Selector menu on the controls bar at the
bottom of the display (left corner).
c.
In the Mode/Trigger dialog box, set Source to
TDR (200 kHz)
.
d.
In the Acq Setup dialog box:
Set Acquisition Mode to
Envelope
.
Under Stop After, check
Condition
and select
Number of
Acquisitions
from the pull down menu.
Set the Number of Acquisitions count to
64
.
DSA8300 Performance Veri
fi
cation
201