![Tektronix 80A00 Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946048.webp)
Performance Tests
80E00 Performance test
Minimum
Incoming
Outgoing
Maximum
zone: from 14 ps
to 150 ps
2
-2%
+25%
zone: from
150 ps to 400 ps
2
-2%
+12%
zone: from
400 ps to 5 ns
2
-2%
+2%
zone: from 5 ns
to 100 ns
2
-2%
+1%
zone: from
100 ns to end
2
-1%
+1%
80E08B
C2 negative polarity
zone: from
-10 ns to -20 ps
1
-1%
+1%
zone: from 14 ps
to 150 ps
2
-2%
+25%
zone: from
150 ps to 400 ps
2
-2%
+12%
zone: from
400 ps to 5 ns
2
-2%
+2%
zone: from 5 ns
to 100 ns
2
-2%
+1%
zone: from
100 ns to end
2
-1%
+1%
80E10B
zone: from
-10 ns to -20 ps
1
-1%
+1%
zone: from 14 ps
to 150 ps
2
-2%
+25%
zone: from
150 ps to 400 ps
2
-2%
+12%
zone: from
400 ps to 5 ns
2
-2%
+2%
zone: from 5 ns
to 100 ns
2
-2%
+1%
C1 positive polarity
zone: from
100 ns to end
2
-1%
+1%