![Tektronix 80A00 Скачать руководство пользователя страница 173](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946173.webp)
DSA8300 Performance Veri
fi
cation
1.
Install the 80E0X module(s) to test starting at the left-most small module
slot as shown in the
fi
gure. Power on the instrument and wait 20 minutes
for warm-up.
2.
Connect the test equipment (See Figure 7.)
3.
Set the synthesized signal generator:
Frequency to
100 MHz (10 ns period)
.
Output to
1600 mV
p-p
(800 mV displayed).
4.
Set the DSA8300 as follows:
a.
Push the
Default Setup
front-panel button and click
Yes
.
b.
Select
C1
from the Waveform Selector menu (left end of controls bar
at bottom of the graticule).
c.
Click
Setup > Mode/Trigger
and select
Direct
for Trigger Source.
d.
Click the
Vert
tab and set the following:
Scale to
100 mV/div
.
Offset to
0
.
e.
Click the
Horz
tab and set the following:
Scale to
5 ns/division
.
Position to the minimum possible value.
Horizontal Reference to
0%
.
Test
Follow this procedure to make the direct trigger level accuracy check:
1.
Click the
Meas
tab.
2.
Select
Meas 1
.
3.
Click the
Source1
button.
4.
Make sure the channel under test is selected in the
Source
tab
5.
Click the
Pulse
button (Signal Type area).
6.
Click
Select Meas > Pulse – Amplitude > Amplitude
.
DSA8300 Performance Veri
fi
cation
155
Содержание 80A00
Страница 2: ......
Страница 6: ......
Страница 16: ...Important safety information x DSA8300 Performance Verification ...