Functional Test Procedures
1.
Set up the instrument:
Click
Utilities > Diagnostics
to open the Diagnostics
dialog box.
2.
Select a diagnostics suite:
a.
Click the
Subsystem
Level tab.
b.
Select all tests (click and hold the
fi
rst entry (Control Proc) and drag down
to highlight all entries as shown above).
c.
Leave Loop and Halt on Failure unchecked (Run area).
3.
Verify that the diagnostic suite passes:
a.
Click
Run
. The diagnostics may take several minutes to complete.
b.
Verify that Pass appears in the dialog box Status column when the
diagnostics complete. If an error number appears as Status, rerun the
diagnostics. If Fail status continues after rerunning compensation and
you have allowed warm up to occur, the module or main instrument may
need service.
c.
Click
Close
to exit.
End of Procedure
Instrument and Module Compensation
Performing a vertical compensation will maximize the accuracy of the automatic
measurements you take. This procedure uses internal routines to optimize the
vertical offset, gain, and linearity.
A complete compensation requires running separate compensations on the
mainframe and the modules.
DSA8300 Performance Veri
fi
cation
3