SVAN 971 User Manual
85
many 1-second integrations as many seconds the
Integration Period
consists
and
stops
the
measurement cycle.
Level -
Type of trigger, that starts the 1-second
measurement/ integration under the condition:
value of the RMS result (
Source
) integrated during
0.5 ms is lower than the threshold value (
Level
). In
other cases, the instrument continues checking the
trigger condition every 0.5 ms. During one
measurement cycle the instrument performs as
many 1-second integrations as many seconds the
Integration Period
consists
and
stops
the
measurement cycle.
Manual
Gr
Type of trigger, that starts the 1-second
measurement/ integration under the condition:
value of the RMS result (
Source
) integrated during
0,5 ms is greater than the threshold level (
Level
)
and the gradient of this Source is greater than the
threshold level (
Gradient
). In other cases, the
instrument continues checking the trigger condition
every 0.5 ms. During one measurement cycle the
instrument performs as many 1-second integrations
as many seconds the
Integration Period
consists
and stops the measurement cycle.
This type of trigger has the same logic as Level +
trigger, but the trigger condition requires also
gradient level to be exceeded.
Manual
Source
Result that is compared with the threshold level
(
Level
) for triggering
– RMS measured in the first
profile: Leq(1).
Manual
Level
Threshold level of the Source for the trigger
condition fulfilment.
Manual
Gradient
Threshold level of the Source gradient for the trigger
condition fulfilment.
Manual
Statistical
Levels
Certain boundary level (
Lnn
) surpassed by the
temporary noise level values in not more than nn
%
of the observation period. In the
Statistical Levels
window, the user can define ten statistical levels,
named from N1 to N10, to be calculated, displayed
and saved in the files together with the main results.
Manual
Appendix D
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...