SVAN 971 User Manual
25
4 MEASUREMENT FUNCTIONS AND CALIBRATION
– Function
In the
Function
section, you can select the measurement
function (
Meas. Funct
) and perform the instrument
calibration (
Calibration
).
To select the
Function
section, press the
<Menu>
key,
select the
Function
position and press
<Enter>
.
<ENT>
4.1 M
EASUREMENT FUNCTIONS OF THE INSTRUMENT
–
M
EASUREMENT
F
UNCTION
The main function of the instrument is measurements of the broadband sound
pressure
level (
Lev. Met.
).
The Sound Level Meter (SLM)
function
meets the standard IEC 61672-1:2013 for Class 1 accuracy. The
instrument can also be used for medium to long-term acoustic monitoring using the huge capacity data
logger in which all measurement results can be stored.
You may also use 1/1 and 1/3 real time octave band analysis and dose meter (dosimeter) options. These
options broaden the main Level Meter functionality of the instrument, because 1/1 and 1/3 analysis as
well as dose meter measurements are performed along with all calculations of the broadband Level Meter
results.
To activate a function, open the
Meas. Funct
list and
select with the
▲
/
▼
key the required function:
Lev. Met.
,
1/1 Oct.
,
1/3 Oct.
,
Dosimeter
or
Run. LEQ
.
<ENT>
Note: Type of measurement function is not displayed on the screen, so the user should
remember the currently selected function!
Note: The 1/1 Octave, 1/3 Octave, and Dosimeter functions are optional and should be
unlocked by entering the activation code in the text editor screen, which is opened after first
attempt to select it. Once unlocked, this option will be ready to use permanently.
Optional functions that broaden the applications of the instrument can be easily installed. These optional
functions can be provided initially by the manufacturer or can be purchased later when required.
Note: It is not possible to change the measurement function during a measurement run. In
this case, the instrument displays for about 3 seconds the text:
“Measurement in Progress”.
To change the function of the instrument the current measurement must be stopped!
4.2
I
NSTRUMENT
’
S CALIBRATION
–
C
ALIBRATION
The instrument is factory calibrated with the supplied microphone for the reference environmental
conditions (see Appendix C). The microphone sensitivity is a function of the temperature, ambient
pressure and humidity, and when the absolute sound pressure level value is required, the absolute
calibration of the measurement channel should be performed. To select the calibration function, open the
Calibration
list.
Due to automatic calibration option the instrument can perform the sound calibration automatically, when
the calibrator is placed over the microphone (switched on or with the auto run function). The calibrator
level is automatically detected, and the calibration measurement is started automatically. Just press
<Enter>
to confirm the calibration results. A sound measurement cannot be in progress while the
automatic calibration is being performed.
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...