APPENDIX A. REMOTE CONTROL
The USB 2.0
interface
is the serial one working with 480 MHz clock which enables one to control remotely
the unit. Its speed is relatively high and it ensures the common usage of USB in all produced nowadays
Personal Computers.
The functions, which are developed in order to control data flow in the serial interfaces, ensure:
- Bi-directional data transmission,
- Remote control of the instrument.
The user, in order to programme the serial interface, has to:
1. send "the function code",
2. send an appropriate data file
or
3. receive a data file.
A.1 I
NPUT
/
O
UTPUT TRANSMISSION TYPES
The following basic input / output transmission types (called functions) are available:
#1
input/output of the control setting codes,
#2
read out of the measurement results in the
SLM
mode,
#3
read out of the measurement results in the
1/1 OCTAVE
analysis or
1/3 OCTAVE
analysis mode,
#4
read out of the data file from the internal Flash-disc or RAM memory,
#5
read out of the statistical analysis results,
#7
special control functions,
#9
writing the data file into the internal flash-disk.
#D
read/write the data file from the external memory (SD-card),
A.2 F
UNCTION
#1
-
I
NPUT
/O
UTPUT OF THE CONTROL SETTING CODES
#1 function enables the user to send the control setting codes to the instrument and read out a file
containing the current control state. A list of the control setting codes is given in Tab. A.1. The format of #1
function is defined as follows:
#1,Xccc,Xccc,(...),Xccc;
or
#1,Xccc,X?,Xccc,(...),X?,Xccc;
where:
X
- the group code,
ccc
- the code value,
X?
- the request to send the current X code setting.
The instrument outputs in this case a control settings file for all requests X? in the following format:
#1,Xccc,Xccc,(...),Xccc;
In order to read out all current control settings the user should send to the device the following sequence
of characters:
#1;
The instrument outputs in this case a file containing all control settings given in Tab. A1 in the format:
#1,Xccc,Xccc,(...),Xccc;
Example:
The instrument sends the following sequence of characters as an answer for the mentioned
above request:
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...