#2,1,T?,R?,V?,P?,L?;
the unit sends out the results of measurement coming from the first profile in predefined, described above,
order:
#2,1,V0,T7,P124.39,R85.86,L(01)100.30,L(10)89.50,L(20)78.60,L(30)68.50,L(40)60.30,L(50)54.00,
L(60)51.00,L(70)46.50,L(80)44.00,L(90)42.40;
Note: All bytes of that transmission are ASCII characters.
.
A.4 F
UNCTION
#3
-
R
EAD
-
OUT OF MEASUREMENT RESULTS IN
1/1 O
CTAVE AND
1/3 O
CTAVE
MODE
#3 function enables one to read out the current measurement results in
1/1 OCTAVE
or
1/3 OCTAVE
modes.
#3 function
format is defined as follows:
#3;
- displayed spectrum
#3,A;
- averaged spectrum
#3,I;
- instantaneous spectrum
#3,M;
- max spectrum
#3,N;
- min spectrum
#3,P;
- peak spectrum
The device responds, sending the last measured spectrum (when the instrument is in STOP state) or
currently measured spectrum (when the instrument is in RUN state) in the following format:
#3;<Status Byte> <LSB of the transmission counter> <MSB of the transmission counter>
<data byte> (...) <data byte>
Status Byte
gives the information about the current state of the instrument.
D7
D6
D5
D4
D3
D2
D1
D0
where:
D7 = 0 means that "overload does not happen",
= 1
means that “overload appeared”,
D5 = 0 means that "spectrum is not averaged ",
= 1 means that "spectrum is averaged ",
D4 = 0 the instantaneous current result (RUN State),
= 1 the final result (STOP State),
D3 = 1 results in
1/3 OCTAVE
mode,
D2 = 1 results in
1/1 OCTAVE
mode,
D6, D1, D0
reserved bits.
Note:
The measurement result is coded in binary form as dB•100 (e.g. 34.5 dB is sent as
binary number 3450).
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...