![Svantek SVAN 971 Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/svantek/svan-971/svan-971_user-manual_1409279032.webp)
SVAN 971 User Manual
32
Level type trigger
The
Level +
trigger starts the 1-second measurement/integration under the
condition: value of the RMS result (
Source
) integrated by 0,5 ms is greater than
the threshold value (
Level
). In other cases, the instrument continues checking the
trigger condition every 0,5 mc.
When the new measurement cycle begins (after pressing the
<Start/Stop>
key or automatically after
stop of the previous measurement cycle) the instrument checks the trigger condition every 0,5 ms and
starts 1-second integration if condition is met.
After 1-second integration, the instrument
repeats the trigger condition checking
every 0,5 ms and starts next 1-second
integration if condition is met. The
instrument does it as many times as many
seconds are within the Integration Period
and stops the measurement cycle.
Therefore,
the
series
of
1-second
measurements may not be continuous,
and the duration of the measurement
cycle
may
be
longer
than
the
Integration Period.
The measurement can be stopped manually at any moment with the
<Start/Stop>
key. Summary Results
are calculated on the base of series of 1-second results measured during each measurement cycle and
saved in the results file.
The
Level -
type trigger starts the 1-second measurement/integration under the
condition: value of the RMS result (
Source
) integrated during 0,5 ms is lower than
the threshold value (
Level
). In other cases, the instrument continues checking the
trigger condition every 0,5 mc.
This is a mirrored trigger to the
Level +
trigger.
Note: When a measurement is waiting for the level trigger, the flashing
“Level trigger” icon superimposes on the „waiting” icon.
/
Gradient type trigger
The
Grad +
trigger starts the 1-second measurement/integration under the
condition: value of the RMS result (
Source
) integrated during 0,5 ms is greater
than the threshold level (
Level
) and the gradient of the Source value is greater
than the gradient threshold level (
Gradient
). In other cases, the instrument
continues checking the trigger condition every 0,5 mc.
This type of trigger has the same logic as the
Level +
trigger, but the trigger condition requires also
gradient level to be exceeded.
Note: When a measurement is waiting for the gradient trigger, the
flashing “trigger” icon superimposes on the „waiting” icon.
Source result
Only one measured result can be used as a source (
Source
) for checking of
trigger conditions in the
Level Meter
mode, namely the output signal from the
LEQ detector coming from the first profile, which is denoted here as
Leq(1)
. This
position does not become active (it is not displayed inversely) and the text stated
here remains unchanged.
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...