SVAN 971 User Manual
– Appendixes
117
APPENDIX B. DATA FILE STRUCTURES
B.1 G
ENERAL STRUCTURE OF THE
SVAN
971
FILES
Each file containing data from the SVAN 971 instrument consists of several groups of words.
In the case of SVAN 971 (the internal file system rev.
1.11
), there are two different types of files containing:
•
the results stored in the file in the instrument’s logger (cf. App. B.2);
•
setup data (cf. App. B.3).
Each file has the following elements:
•
SvanPC file header (cf. Tab. B.1.1)
•
file header (cf. Tab. B.1.2);
•
unit and internal software specification (cf. Tab. B.1.3);
•
calibration settings (cf. Tab. B.1.4)
•
user’s text (a header) stored together with the measurement data (cf. Tab. B.1.5);
•
parameters and global settings, common for all profiles (cf. Tab. B.1.6);
•
parameters for measurement trigger (cf. Tab. B.1.7);
•
parameters for logger trigger (cf. Tab. B.1.8);
•
parameters for Time-domain signal recording (cf. Tab. B.1.9);
•
parameters for Wave-file recording (cf. Tab. B.1.10);
•
special settings for profiles (cf. Tab. B.1.12);
•
display settings of the main results (cf. Tab. B.1.13)
•
header of the statistical analysis (cf. Tab. B.1.14);
•
header of the file from the logger (cf. Tab. B.1.15)
•
contents of the file from the logger (cf. Tab. B.1.16)
Other elements of the file structure are not obligatory for each file type stated above.
They depend on the file type (
SLM
,
DOSE METER
, file from the logger) and on the setting of the
FULL STAT.
These elements are as follows:
•
Header of the Summary Results Record (saved in Summary Results Record) (cf. Table B.1.17)
•
main results (saved in Summary Results Record)
(cf. Tab. B.1.18_SLM, B.1.18_DM )
•
statistical levels (saved in Summary Results Record)
(cf. Tab. B.1.19)
•
1/1 OCTAVE analysis results (saved in Summary Results Record)
(cf. Tab. B.1.20)
•
1/3 OCTAVE analysis results (saved in Summary Results Record)
(cf. Tab. B.1.21)
•
results of the statistical analysis (saved in Summary Results Record) (cf. Tab. B.1.22);
•
settings of the instrument saved in the setup file (cf. Tab. B.1.23);
•
file-end-marker (cf. Tab. B.1.24);
Below, all file structure groups are described separately in Tab. B.1.1
– Tab. B.1.23. The format used in
the columns, named
Comment
with the square parenthesis ( [xx, yy] ), means the contents
of the word with;
xx
is the most significant byte (MSB) and
yy
the lowest significant byte (LSB)
of the word. The format 0xnnnn means that the nnnn is four-digit number in hexadecimal form.
Table B.1.1. SvanPC file header
Word
number Name
Comment
0..2
“SvanPC”
reserved
3
26
reserved
4
32
reserved
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...