4.2.4 Automatic calibration
– Auto Calibration
Position
Auto Cal.
enables the user to perform automatic
calibration when the sound calibrator is attached. In this
case, the window
Calibration
by measurement will
appear automatically. If
Auto Cal.
is switched off, the
user should enter this window through the
Menu
.
<ENT>
Automatic calibration feature was implemented to make calibration as easy as possible and to allow the
user to perform a calibration of the SV 977A with minimum steps.
If the automatic calibration is switched on, the instrument, when it doesn’t perform the measurement,
periodically compares the measured signal level (Running SPL for 1 second) with the reference
calibration level and starts the calib
ration measurement if the stable SPL result is within ±5dB of the
calibration level.
To perform the automatic microphone calibration, follow next steps:
1. Switch on the instrument.
2. Attach the SV 35A (or equivalent 114 dB/1000 Hz) calibrator to the microphone and switch it on (if
the used calibrator doesn’t have automatic switch-on feature).
Generated by the calibrator sound pressure level starts the Automatic Calibration
process if the difference between the
Calibration
Level
value set up in the
Calibration
screen and the measured calibrator SPL level is in the range
±5 dB
.
During the calibration measurement, the level of the measured calibration signal
will be displayed. If the maximal difference between three 1-second consecutive
Leq results is less than
0.05dB
, the calibration measurement will be finished. The
measurement can be always stopped by the
<Start/Stop>
key.
After calibration measurement stop, the
Old
and
New Factor
(difference
between the Calibration Level and the Calibration Measurement, calculated in
dB) are displayed. The user will be proposed to save the new calibration factor
by pressing
<Enter>
(
Enter to Confirm
), or reject it by pressing
<Esc>
. In both
cases the instrument exits the
Calibration
screen.
3. Detach the calibrator from the microphone and leave the
Auto Calibration
screen.
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...