SVAN 971 User Manual
35
Switching on the
Logger
(
On
) activates other positions in the
Logging
list, which
enable the user to save selected results from the three profiles, spectra with the
defined interval selected in the
Logger Step
position (
Logger Results
), program
the
Logger Trigger
and recording of the time signal (
Event Recording
).
Note: If Logger is Off, result files are not created, and measurement results (both summary
and logger) are not saved!
Splitting of the logger file
The
Log.
Split
position enables splitting the logger data registration into separate files. If
Log.
Split
parameter is
Off
the registration of measurement results will be continuously made in one logger file with
the name defined in the
Logger Name
position.
In other cases, the registration will be made in separate files and the registration
in the new file will start: after integration period time (
Integr. Per
), or every quarter
of the RTC (
Sync. to 15m
), or every half an hour of the RTC (
Sync. to 30m
), or
every hour of the RTC (
Sync. to 1h
), or at specified by the user times
(
Spec. Time
). Whenever the split time is achieved the logger file is closed and
the new file with the increased by one number is opened for subsequent
measurement data.
Note: In the Simple interface mode, the Logger Split parameter is hidden, but the instrument
will use settings previously defined in the Advanced mode or default settings (Off)
.
If
Spec. Time
is selected in the
Log.
Split
position, then
six additional positions (
Split Time1
,
Split Time2
,
Split Time3
,
Split Time4
,
Split Time5
and
Split Time6
)
appear in the list. The position is switched off if
Off
was
selected. Pressing the
►
key changes the
Off
to the time
format value. Further use of the
►
key enables setting
the desired time of the day when splitting should occur.
►
The
Logger Step
defines the step for logger results logging in a file. It can be set
from
100ms
to
1h
. Its value by default is set to
1s
.
In the
Simple
instrument interface mode, this parameter is hidden, but the
instrument will use previously defined or default value.
Note: In the Simple interface mode, the Logger Step parameter is hidden, but the instrument
will use settings previously defined in the Advanced mode or default settings (1s)
.
Logger file name
The
Logger Name
position enables defining the logger
file name, which consists of a prefix and a number. The
default logger file prefix is
L
. The name can be up to eight
characters long. After pressing the
◄
/
►
key, the special
window with text editor function is opened for the file
name editing.
►
The edited name is accepted and saved after pressing
the
<Enter>
key. The special warning is displayed in case
the file with the same name already exists in the memory.
The instrument
informs with the message “Incorrect File
Name” and waits for the
<Enter>
key to be pressed.
If the name is new the instrument changes the
Logger Name
in the
Logger Setup
list.
<ENT>
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...