![Svantek SVAN 971 Скачать руководство пользователя страница 218](http://html1.mh-extra.com/html/svantek/svan-971/svan-971_user-manual_1409279218.webp)
SVAN 971 User Manual
– Appendixes
218
PTC
The PTC result (Peak Threshold Counter)
– the number of the overpasses of the Threshold
Level by Lpeak result. This result is incremented in 100 ms intervals.
PTP
The
PTP
result is the
PTC
result expressed
in percent.
ULT
Upper Limit Time - time
that SPL exceeded the “ULT Threshold Level” set during
configuration.
TWA
The Time Weighted Average is the average
A-weighted sound level for a nominal 8-hour
workday with Time Weighting S and
Exchange Rate 5.
TWA
is usually measured
with A-weighting and Slow response detector
type. TWA is calculated from the measured
LAV (taking Threshold Level into account)
and a Reference time of 8 h. Mainly used in
the USA for assessing the noise exposure for
a worker during a workday.
•
Sound
levels
at
or
above
the
THRESHOLD LEVEL
are averaged into
the
calculations
relating
to
noise
exposure.
TWA
is calculated with no
threshold level, or with threshold level
(typically 80dB or 90dB)
•
In case the time period is below 8 hours,
the
TWA
is less than the
LAV
In case the time period is more than 8
hours, the
TWA
is greater than the
LAV
PrTWA
The Projected Time Weighted Average is calculated from the measured
LAV
(taking
THRESHOLD LEVEL
into account) and the exposure time.
Lc-a
The C-A measurement is an Leq that
enhances the low-frequency components of
the sound signal. It is the result of subtracting
an A-weighted LAeq from a simultaneously
collected C-weighted Leq
Lc-a = LCeq - LAeq
D.4
D
EFINITIONS AND FORMULAS OF THE ADDITIONAL
R
UNNING
LEQ
FUNCTION RESULTS
LR15
15-minutes running Leq is the rolling (sliding)
Leq window for the last 15 minutes of
measurement (900 seconds) moving with 1
second step
T
900
T
2
0
dt
p
t
r
s
900
1
log
10
15
LR
Note: If the current time period of the measurement T is less then 15 minutes then LR15 result is
undefined.
LR60
60-minutes running Leq is the rolling (sliding)
Leq window for the last 60 minutes of
measurement (3600 seconds) moving with 1
second step
T
3600
T
2
0
dt
p
t
r
s
3600
1
log
10
60
LR
Note: If the current time period of the measurement T is less then 60 minutes then LR60 result is
undefined.
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...