SVAN 971 User Manual
– Appendixes
149
Maximum peak voltage
20 V Peak-Peak (Maximum peak voltage of input sinusoidal
signal, which can be lead to the SLM without destruction the
meter)
Warm-up time
1 min. (for 0.1 dB accuracy)
Nominal delay
less than 3 seconds
(between operating of the "Reset- Button"
and beginning of a new measurement)
Typical stabilization time
after change in
environmental conditions by 20
C
1 hour
Nominal delay
between operating of the
"Reset- Button" and beginning of a new
measurement
< 3 sec
Time shift after completion
of a
measurement, before a measurement is
shown
< 1 sec
Note: When the instruments are moved from a warm environment with high humidity,
to a colder environment, care should be taken not to produce condensation inside
the instruments. In this case, much longer stabilization periods may be necessary.
Environmental, electrostatic
and radio frequency
criteria
Effect of humidity
< 0.5 dB (for 30%<RH<90% at 40
C and 1000 Hz)
Effect of magnetic field
< 15 dB (A) or < 25 dB (Z) (for 80 A/m and 50 Hz)
Effect of radio frequency fields
< +/-0.5 dB @ 74 dB and 10V/m
electromagnetic field
The greatest susceptibility (the least immunity) is achieved when in the SLM the
Z
filter and time weighting
F
are selected, and the SPL measurements are considered.
The greatest susceptibility is achieved when the SLM is placed parallel to the radio frequency field.
In addition, if there is an extension cable, the greatest susceptibility is achieved when the SLM and cable is
placed along field and the cable is coil as solenoid.
Effect of electrostatic discharge
meets requirements of IEC 61672-1:2013
During electrostatic discharge, the influence of the displayed results could be observed.
No changes in instrument operation state, configuration or stored data corruption were found out.
Effect of ambient pressure
< 0.01 dB/kPa
Effect of temperature
< 0.5 dB (from -10
C to + 50
C)
Operating range
from -10
C to + 50
C
Storage
from -20
C to + 60
C
Effect of Vibration
1.
For mechanical vibrations with an acceleration of 1 m/s² vertically to the membrane of the microphone for
the frequencies 31,5 Hz, 63 Hz, 125 Hz, 250 Hz, 500 Hz, 630 Hz, 800 Hz and 1000 Hz increases the low
level of the linear operation range to 70dB for the frequency weighting “A”.
2.
For mechanical vibrations with an acceleration of 1 m/s² parallel to the membrane of the microphone for
the frequencies 31,5 Hz, 63 Hz, 125 Hz, 250 Hz, 500 Hz, 630 Hz, 800 Hz and 1000 Hz increases the low
level of the linear operation range to 65dB for the frequency weighting "A".
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...