SVAN 971 User Manual
– Appendixes
103
and 3 when the under-range took place during the last measurement period and it lasted in the last
second of the measurement);
V
overload flag (ccc equals to 0 or 1);
T
time of the measurement (ccc
– value in seconds);
x
start date of the measurement in format
dd/mm/yyyy
(
dd
– day,
mm
– month,
yyyy
- year)
t
start time of the measurement in format
hh/mm/ss
(
hh
– hour,
mm
– minute,
ss
- second)
P
Lpeak
value (ccc
– the value in dB);
M
Lmax
value (ccc
– the value in dB);
N
Lmin
value (ccc
– the value in dB);
S
L
result (ccc
– the value in dB);
D
DOSE
result (ccc
– the value in %);
d
D_8h
result (ccc
– the value in %);
p
PrDOSE
result (ccc
– the value in %);
A
LAV
result (ccc
– the value in dB);
R
Leq
result (ccc
– the value in dB);
U
LE
result (ccc
– the value in dB);
u
SEL8
result (ccc
– the value in dB);
E
E
result (ccc
– the value in Pa
2
h);
e
E_8h
result (ccc
– the value in Pa
2
h);
I(nn)
LEPd
result (ccc
– the value in dB, nn – the value of Exposure Time in minutes);
J
PSEL
result (ccc
– the value in dB);
Y
Ltm3
result (ccc
– the value in dB);
Z
Ltm5
result (ccc
– the value in dB);
L(nn)
value L of the
nn statistics (ccc
– the value in dB);
C
PTC
result (ccc
– the counter value);
c
PTP
result (ccc
– the value in %);
l
ULT
result (ccc
– value in seconds);
W
TWA
result (ccc
– the value in dB);
w
PrTWA
result (ccc
– the value in dB);
a
Lc-a
result (ccc
– the value in dB);
The exemplary results of the instrument’s response after sending to it the following sequence
of characters:
#2,1;
coming from the first profile are given below:
a) for the case of the
SLM
mode:
#2,1,x17/03/2014,t13:44:28,v0,V0,T10,P79.97,M52.92,N38.50,S46.35,R43.91,U53.91,B(1)43.91,I(
480)43.92,Y50.67,Z51.15,L(01)55.00,L(10)45.60,L(20)44.30,L(30)42.80,L(40)41.50,L(50)40.80,L(6
0)40.40,L(70)40.00,L(80)39.50,L(90)39.00;
b) and for the case of the
DOSE METER
mode:
#2,1,x17/03/2014,t13:48:36,v0,V0,T7,P124.39,M99.26,N41.54,S42.05,D0,d389,p389,A85.86,R85.
86,U94.31,u130.45,E0.00,e1.23,I(480)85.87,J49.72,Y95.62,Z99.22,L(01)100.30,L(10)89.50,L(20)7
8.60,L(30)68.50,L(40)60.30,L(50)54.00,L(60)51.00,L(70)46.50,L(80)44.00,L(90)42.40,C4,c6,l0,W4
9.72,w85.87,a-0.55;
Note: The presented above order of the measurement results sent out by the instrument does
not depend about the characters sent to the unit.
.
Example:
After sending to the instrument the string:
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...