SVAN 971 User Manual
64
The averaging of results for each spectrum band is performed during the
Integration Period
and is repeated the
Repetition Cycles
times.
Both parameters are defined in the
General Settings
list.
11.2.1 Selecting the measurement range for the 1/1 and 1/3 octave analysis - Range
For the
1/1 Octave
or
1/3 Octave
functions you can
select the input ranges specified in Appendix C, named
as
Low
and
High
.
The selection of the input range is made in the
Range
window of the
Measurement
list.
►
The above screens were made with calibration factor
equal to zero. The calibration factor is always added to
the upper range level
– see example.
=>
11.2.1 Logging of 1/1 and 1/3 octave spectra
– Logging
Spectra are always logged together with Summary results in a logger file with
Integration Period
step. The first condition should be fulfilled, namely the
Logger
must be switched on (
path: <Menu> / Measurement / Logging / Logger Setup /
Logger: On).
The
Leq
and
Lpeak
results from 1/1 Octave or 1/1 Octave analysis can also be
saved in the logger file with the step defined by the
Logger Step
parameter (
path:
<Menu> / Measurement / Logging / Logger Setup
). The enabling of spectrum
saving in the logger file is made by checking the
Peak
Spectrum
or
Leq Spectrum
position with the
◄
/
►
key.
11.2.2 Setting up the parameters of 1/1 and 1/3 octave analysis - Spectrum
For active
1/1 Octave
or
1/3 Octave
functions the
additional position (
Spectrum
) appears on the
Measurement
list.
The
Spectrum
position enables selecting the pre-
weighting broadband frequency filter and LEQ detector
for the octave or third octave analysis.
<ENT>
Following weighting filters are available for the 1/1 and 1/3 octave
analysis:
•
A
class 1 according to the IEC 651 and IEC 61672-1:2013 standards,
•
C
class 1 according to the IEC 651 and IEC 61672-1:2013 standards,
•
Z
class 1 according to the IEC 61672-1:2013 standard,
•
B
class 1 according to the IEC 651 standard.
Filter characteristics are given in Appendix C.
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...