SVAN 971 User Manual
68
12 DOSIMETER
The instrument operates as a
Dosimeter
in a very similar way to the
Level Meter
and, in addition to
SLM results, measures also basic dose parameters.
Below additional settings related with the
Dosimeter
function are described.
12.1 S
ELECTION OF THE
D
OSIMETER FUNCTION
To select the
Dosimeter
function, enter the
Function
section, select the
Meas.
Funct
position and press the
<Enter>
key. In the
Meas. Funct
window, select the
Dosimeter
function and press the
<Enter>
key.
<ENT>
12.2
S
ETTING GENERAL PARAMETERS
–
G
ENERAL
S
ETTINGS
Most general settings of the
Dosimeter
function are similar to the
Level Meter
function (see chapter 5.1).
In addition,
Dosimeter
has a programmable five automatic pauses.
Programable automatic pauses
Automatic pause(s) can be switched off (
Off
) or can be programmed based on
absolute time (
On
).
If
Pause
is
On
, two additional positions appear which
enable setting time for pause start (
Pause
Begin
) and
time for pause end (
Pause End
).
▼
Note: In the Simple interface mode, the Pause parameters are hidden, but the instrument
will use settings previously defined in the Advanced mode or default settings (Off)
.
12.3 S
ETTING PROFILE PARAMETERS
–
P
ROFILE
X
Parameters of three profiles can be set in the
Profile x
lists of parameters.
The following parameters can be programmed
independently for each profile: weighting filter
(
Filter
),
peak filter (
Filter Peak
) and LEQ detector type
(
Detector
), criterion level (
Crit. Level
), threshold level
(Thr. Level
), exchange rate (
Exch. Rate)
,
ULT Thresh.
and
PTC Thresh.
.
<ENT>
Weighting filter selection
•
Z
class 1 according to IEC 61672-1:2013 standard,
•
A
class 1 according to IEC 651 and IEC 61672-1:2013 standards,
•
C
class 1 according to IEC 651 and IEC 61672-1:2013 standards,
•
B
class 1 according to IEC 651 standard.
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...