SVAN 971 User Manual
26
The
Calibration
list comprises positions enabling:
calibration with the use of the sound calibrator
(
By Measure.
),
checking
previous
calibrations
(
Last Cal.
), adding current calibration results to the
logger file (
Post Cal.
).
<ENT>
In the
Advanced
user interface, there is additional position in the
Calibration
list
-
Auto Cal.
, which allows you to switch on the auto calibration function.
Note: It is advised to perform calibration of the instrument each time before the
measurements begin. A single calibration at the start of each day is usually sufficient for most
regulations.
Note: The calibration factor is always added to measurement results and measurement range
limits of the Lev. Met., 1/1 Oct., 1/3 Oct. and Dosimeter functions.
Note: The recommended factory calibration interval is 12 months for instruments to be
confident in their continuing accuracy and compliance with the international specifications.
Please contact your local Svantek distributor for further details.
4.2.1 Calibration - By Measurement
To calibrate the instrument:
1. Set the calibration level (
Cal. Level
) appropriate to the used calibrator. The
default level is 114 dB at 1000 Hz. Remember to change this level if using an
alternative reference sound signal source.
2. Attach the sound calibrator (SV 35A or equivalent 114 dB/1000 Hz)
carefully
over the microphone of the instrument.
Note: It is also possible to use an electro-mechanical pistonphone, which generates the
signal (ca 124 dB) or different type of sound calibrator
dedicated for ½” microphones. In any
case, before starting the calibration measurement, you should set the level of the signal
generated by the given calibrator (Cal. Level position), which is stated in the calibrat
or’s
certificate (the value of the Cal. Level set by the manufacturer of SVAN 971 is equal to
114 dB). It is also necessary to switch the instrument Range to the High level.
3. Switch on the calibrator
(if the used calibrator doesn’t have auto run function)
and wait ca 30 seconds for the tone to stabilise before starting the calibration
measurement.
4. Start the calibration measurement by pressing the
<Start/Stop>
key.
The calibration delay time is set to 3 seconds. While waiting for the start of the
measurement the
Delay
is counting down on the display.
During the calibration measurement, the level of the measured calibration signal
is displayed. If the maximal difference between three 1-second consecutive
results (Leq(C)) is less than
0.05dB
, the calibration measurement will be stopped,
and the calibration factor will be calculated. The measurement can be always
stopped by the
<Start/Stop>
key.
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...