5.5.4 Event recording settings
– Event Recording
The
Event Rec.
position enables activating and setting
parameters of a waveform signal recording in the logger
file. Event records are placed in the logger file together
with
Summary Results
and
Logger Results
. All records
are synchronized in time that enables synchronous post
measurement processing of all measured data.
In the
Simple
instrument interface mode, the
Event Rec.
position doesn’t appear in the
Logging
list.
<ENT>
Note: In the Simple interface mode, the Event Recording position is hidden, but the
instrument will use settings previously defined in the Advanced mode or default settings
(Recording: Off)
.
Note: The Event Recording function is optional and should be unlocked by entering the
activation code in the text editor screen, opened by the ► key. Once unlocked this option will
be ready to use permanently.
The
Recording
position, if it is not
Off
, defines the way a
signal recording should be done, continuously during the
measurement (
Continuous
) or on the trigger:
Slope +
,
Slope -
,
Level +
,
Level -
,
Gr
,
Trig.manual
or
Integr. Period
.
►..►
Slope type trigger
The
Slope +
trigger starts an event recording under the condition: rising value of
the RMS result (
Source
) averaged by 0,5 ms passes above the threshold level
(
Level
).
After pressing the
<Start>
key the instrument checks the trigger condition with intervals, defined by the
Tr. Period
parameter, and if condition is met starts the event recording. The recording lasts minimum
time, defined by the
Rec. Time
parameter, and during this time the instrument continues to check the
trigger condition (provided that the
Tr. Period
is shorter than the
Rec. Time
). If next trigger condition is
met during the
Rec. Time
the instrument triggers recording again, so it will be continued from this
moment by additional
Rec. Time
and so on. If during next recording time there are no triggers, the
recording will be stopped after the last trigger plus
Rec. Time
. Assuming, that after first recording the
trigger conditions continue to be checked, and new event recording may start during the same
measurement time.
The example shows that between
measurement start and stop two records
were created. The first record is equal to
the
Rec. Time
, because during this period
no second trigger condition has been met.
The second recording is stopped by the
measurement stop and this record can be
shorter than the
Rec. Time
.
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...