![Svantek SVAN 971 Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/svantek/svan-971/svan-971_user-manual_1409279014.webp)
SVAN 971 User Manual
14
2 MANUAL CONTROL OF THE INSTRUMENT
Control of the instrument has been developed in a fully interactive manner. The user can operate the
instrument by selecting the appropriate position from the screen
Menu
list. Thanks to that, the number
of the control keys of the instrument has been reduced to eight for ease of use and convenience.
2.1 C
ONTROL KEYS ON THE FRONT PANEL
The following control keys are located on the front panel of the
instrument:
•
<ESC>
,
(<P/S>)
•
<Enter>
,
(<Menu>)
•
▲
,
◄
,
►
,
▼
•
<Shift>
•
<Start/Stop>
The action given in (...) brackets denotes the second key function
which is available after pressing it in conjunction (or in sequence)
with the
<Shift>
key.
<Shift>
The second function of a key (
<P/S>
,
<Menu>
) can be used when the
<Shift>
key is
pressed together with
<Enter>
,
<ESC>
or some other keys. This key can be used in
two different modes, which can be configured in the
Keyboard
list (
path: <Menu> /
Instrument / Keyboard
):
•
like in a computer keyboard, when both
<Shift>
and the second key must be
pressed simultaneously (
Direct
mode);
•
like in a smartphone keyboard, when the first
<Shift>
key should be pressed
and released and then the second key pressed (
2nd Function
mode).
Note: Simultaneous pressing of the <Shift> and <Start/Stop> keys turning the instrument on
or off.
<Start/Stop>
This key allows you to start and stop the measurement process.
<Enter>
This key allows you to open the selected position on the Menu list, to confirm selected
settings or to switch the views of the result presentation modes. Some additional
functions of this key will be described in the following chapters of this manual.
(<Menu>)
This key (pressed together with the
<Shift>
) allows you to enter the main
Menu
containing seven sections:
Function
,
Measurement
,
Display
,
File
,
Instrument
,
Aux. Setup
and
Report
. Each section consists positions, which opens screens with list
of positions or configuration parameters. These sections will be described in detail in
the following chapters of the manual. Double pressing of the
<Menu>
key opens the list
containing the last earlier opened eight lists of parameters. It often speeds up the control
of the instrument as you have faster access to the frequently used lists of parameters
for easy navigation.
<ESC>
This key closes lists of parameters or other screens and return to the upper list of the
menu. It acts in an opposite way to the
<Enter>
key. When the window is closed after
pressing the
<ESC>
key, any changes just made are ignored.
(<P/S>)
This key allows you to pause or break the measurement process temporarily. If there is
no current running measurement in progress this key opens the Setup Manager menu.
◄
/
►
These keys allow you, in particular, to:
•
select the column in a multi-column parameter list;
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...