SVAN 971 User Manual
42
Note: In the Simple interface mode, Compensation Filter position is hidden, but the
instrument will use settings previously defined in the Advanced mode or default settings
(Microphone: On; Diff. Field: Off; Windscreen: Off; Outdoor: Off)
.
5.8 S
TATISTICAL LEVELS SETTINGS
–
S
TATISTICAL
L
EVELS
In the
Stat. Lev.
list of parameters, the you can define ten
statistical levels, named from
N1
to
N10
, to be calculated,
displayed and saved in a file together with the Summary
results.
Default statistical levels have following settings:
1
,
10
,
20
,
30
,
40
,
50
,
60
,
70
,
80
and
90
. All values should be within
the integer range [1, 99]. Each value can be set
independently from others.
<ENT>
5.9 P
ROGRAMMING THE INSTRUMENT
’
S INTERNAL TIMER
–
T
IMER
The
Timer
function is used to programme the automatic start-up of the measurement at a given time and
day of a week and with the parameters set in the
Measurement
section.
The
Timer
position enables programming the internal
real-time clock to act as a delayed start and stop timer.
The instrument will be switched on by itself at the
programmed time and will perform the measurement with
the same settings that were used before the instrument
was turned off.
<ENT>
Note: In the Simple interface mode, the Timer position is hidden, but the instrument will use
settings previously defined in the Advanced mode or default settings (Timer: Off)
.
Setting hour and day of the measurement’s start
The
Start (hh:mm)
and
Stop (hh:mm)
positions
determine the time for the measurement to start and to
stop automatically.
►
In the positions:
Monday
,
Tuesday
, …,
Sunday
; you can select days in a week
when measurements should start.
The timer can be programmed for
Max days
ahead (
up
to 100) or without
limitation (
Inf
) and during these days, the instrument refers to the time of the
R
eal
T
ime
C
lock (
RTC
). Make sure that the real-time clock settings are correct before
using the timer.
Note: Make sure that there is sufficient internal batteries power available for the instrument
to carry out the required measurements when it wakes up.
5.10 E
XAMPLE OF TIMER EXECUTION
Let us assume that you wish to switch on the measurement on Monday at 22:00,
to measure noise level for 10 minutes and save results in the fileat s with names
R1, R2, R3 etc. as long as they are created
To do this configure the
Timer
function as on the attached screen and to set the
measurement parameters (
path: <Menu> / Measurement / General Settings
) and
the file name (
path: <Menu> / Measurement / Logging / Logger Setup
).
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...