Section 20 - Page 18
AutoDyn manual 1174 Sect20.PM6.5, 05/1999
20.4.1.2.
Using
Step 1
: Prior to updating the system software:
a) Install any EPROM’s that may have been provided,
b) ensure that the system is powered on and functioning
correctly within the network for this computer, and
c) ensure that that the dynamometer is not in use and no tests
or calibrations are in progress.
Step 2
: Select
Tools|Update Remote System Software
from the
WinDyn main menu.
Step 3
: After reviewing the initial information on the “SuperFlow
Automated System Update Program” window, click ‘
Proceed
’
to proceed.
Step 4
: A warning message will appear. Please review and click ‘
OK
’
when ready to proceed.
Step 5
: This procedure will now interrogate your network to locate
and identify any SuperFlow systems. When completed, a new
window will appear instructing you to select the system to
update. Click ‘
OK
’ after selecting the system. If your system
does not appear, click ‘
Cancel
’ and try to correct the problem
before running the procedure.
Step 6
: The update procedure will automatically proceed. No further
user intervention is required. Please wait until the entire
procedure is complete (about 1 minute). The “SuperFlow
Automated System Update Program” window will indicate the
current action being performed. A status bar will be updated
to indicate an approximate percent completion of the entire
procedure. Please do not perform any other WinDyn function
while the update is in progress.
Step 7
: Wait until the procedure indicates “Update complete...”. The
status bar will have been removed and the ‘
Finish
’ button
enabled. Click the ‘
Finish
’ button to exit procedure.
This system has now been updated and is ready to use.
20.4.1.3.
Functional Description...
This procedure first interrogates your network to locate and
identify any SuperFlow systems. You will then be prompted to
select the system to update from a list. After the selection is
made, this procedure will first retrieve the current calibration
information for the system. Following this, the software for your
system will be downloaded into each node. These nodes may
be a sensor box, handheld, fuel system, table top unit, and other
devices as appropriate for your NGE system type. Following the
download, each node is reinitialized and communications will
be reestablished with WinDyn. After this initialization, the
previously saved calibration information will be returned to your
system. The system will then be ready to use.
Содержание AutoDyn
Страница 4: ...AutoDyn Operators Manual Section 1 Page 4 AutoDyn Cover fm...
Страница 5: ...1 General Information...
Страница 10: ...S Y S T E M O V E R V I E W Section 2 Page 1 2 System Overview...
Страница 24: ...P A C K A G I N G A N D H A N D L I N G Section 3 page 1 Packaging and Handling 3...
Страница 29: ...T E S T A R E A R E Q U I R E M E N T S Section 4 page 1 4 Test Area Requirements...
Страница 40: ...I N S T A L L A T I O N Section 5 Page 1 5 Installation...
Страница 55: ...6 System Operation...
Страница 60: ......
Страница 78: ...H A N D H E L D C O N T R O L L E R Section 7 Page 1 7 Handheld Controller...
Страница 104: ...8 Theory Of Testing...
Страница 114: ...Theory Of Testing Volume 2 9 10 04 Section 8 Page 13 Figure 8 1 Frontal Area Coefficient of Drag US...
Страница 115: ...AutoDyn Operators Manual Section 8 Page 14 Theory Of Testing fm Figure 8 2 Frontal Area Coefficient of Drag Metric...
Страница 123: ...9 Description of the Configuration File...
Страница 172: ...20 Service and Calibration...
Страница 178: ...S E R V I C E A N D C A L I B R A T I O N Section 20 page 7 20 2 Maintenance record...
Страница 193: ...A General Testing Information Appendix...
Страница 205: ...B Fundamental Concepts to Obtaining Superior Repeatability Appendix...