Description of the Configuration File
Volume 2, Rev 2 6/28/04
Section 9 - Page 9
Channels 21 to 28 - EXHAUST TEMPERATURES 1 THROUGH 8
Channels 21 through 28 are used for measuring exhaust gas temperature
thermocouples. They may be used for other temperature measurements
if desired. These channels are normally set up for Type K thermocouples.
No other type of thermocouple can be used without a hardware
modification. The maximum temperature input is 2000
o
F
[1100
o
C]
. These
channels require no zeroing and are set with a filter rate of 5.
Channels 29 to 36 - ANALOG VOLTAGES 1-8
Channels 29 through 36 are used for measuring DC voltage inputs. They
are currently defined for 0-10 vdc inputs on channels 29-35. Channel 36
is normally defined for 0-20 vdc. Each channel may be configured via
hardware modifications for 0-1, 0-5, 0-10, or 0-20 vdc. These channels
are set for zeroing by command and are set with a filter rate of 5. The
channel is set for a coefficient of 1.000 and does not require calibration.
Channels 29-35:
0.00
10.00 vdc
0.00
10.00 vdc
Channel 36
0.00
20.00 vdc
0.00
20.00 vdc
Channels 37 through 49 - UNUSED
Channels 37 through 49 are not currently used in the standard version of
the dynamometer. Eight expansion sensors may be added in this channel
slot with optional, additional hardware. The additional inputs could be
thermocouples, pressures, or analog voltages. These may be renamed
to fit specific functions by using the Configuration Editor Program. They
may also be used for additional calculated channel definitions.
Channel 61 - Oil P
Channel 61 is for measuring oil pressure. The transducer has a range of
0-300 psi
[2000kPa]
. It may be recalibrated if required, through the
calibration menu, or through changes in the Configuration File. Filtering is
set to 5 and zeroing is by command.
0.000 3.636 volts
0.000 150 psi
[1034 kPa]
Channel 63 - Man. P
Channel 63 is used to measure the manifold inlet pressure or vacuum in
inches of mercury. It has a range of -30 to +173 inches of mercury
[-103
to +586 kPa]
. Filtering is set to 5 and zeroing is by command.
0.000 3.636 volts
0.000 407 in.Hg
[1379 kPa]
Содержание AutoDyn
Страница 4: ...AutoDyn Operators Manual Section 1 Page 4 AutoDyn Cover fm...
Страница 5: ...1 General Information...
Страница 10: ...S Y S T E M O V E R V I E W Section 2 Page 1 2 System Overview...
Страница 24: ...P A C K A G I N G A N D H A N D L I N G Section 3 page 1 Packaging and Handling 3...
Страница 29: ...T E S T A R E A R E Q U I R E M E N T S Section 4 page 1 4 Test Area Requirements...
Страница 40: ...I N S T A L L A T I O N Section 5 Page 1 5 Installation...
Страница 55: ...6 System Operation...
Страница 60: ......
Страница 78: ...H A N D H E L D C O N T R O L L E R Section 7 Page 1 7 Handheld Controller...
Страница 104: ...8 Theory Of Testing...
Страница 114: ...Theory Of Testing Volume 2 9 10 04 Section 8 Page 13 Figure 8 1 Frontal Area Coefficient of Drag US...
Страница 115: ...AutoDyn Operators Manual Section 8 Page 14 Theory Of Testing fm Figure 8 2 Frontal Area Coefficient of Drag Metric...
Страница 123: ...9 Description of the Configuration File...
Страница 172: ...20 Service and Calibration...
Страница 178: ...S E R V I C E A N D C A L I B R A T I O N Section 20 page 7 20 2 Maintenance record...
Страница 193: ...A General Testing Information Appendix...
Страница 205: ...B Fundamental Concepts to Obtaining Superior Repeatability Appendix...