Return To Table of Contents
www.stryker.com
2141-209-002 REV H
99
Service Information
BATTERY REMOVAL AND REPLACEMENT - (LITTER)
Tools Required:
• Phillips Screwdriver #2
• 5/16” Nutdriver
• 5/16” Combination Wrench
Procedure:
Note
Batteries should always be replaced in pairs.
1. Plug bed into wall outlet.
2. Set the brakes using one of the brake control locations or by using the manual brake pedal.
3. Raise the bed to the full up position.
4. Remove the mattress assembly or fold back to expose the foot section.
5. Using a Phillips screwdriver, remove the two screws securing the electrical cover and remove the cover
(reference Figure 1 above, item A and B).
Note: Use caution as the cover is large and heavy.
6. Unplug the bed from the wall outlet and turn battery switch “OFF” (0). The bed should now have no power.
7. Unplug J12 connector from the CPU/Power board.
8. Lift up on both batteries and stand them upright (reference Figure 2 above).
9. Using a 5/16” nutdriver and a 5/16” combination wrench, unbolt the wires from the battery posts.
10. Reverse the steps to install new batteries.
Note: New battery information should be facing the right side when installing and upright when installed.
11. Test all bed functionality while plugged into wall outlet and when unplugged prior to putting back into service.
A
B
Electrical Cover
Batteries
Figure 2
Figure 1