Return To Table of Contents
www.stryker.com
2141-209-002 REV H
59
Error Codes
Touch Screen Error Messages Name Definition
Cmd WO safe from room
Communication board has a network communication error (Check network
connections).
Cmd WO safe from room
Communication board has a network communication error (Check network
connections).
GPIO Failure Init
Initialization of the PCA9555 (GPIO expansion chip) failed (Replace control
board).
GPIO Failure Read
Reading from the PCA9555 (GPIO expansion chip) failed (Replace control
board).
GPIO Failure Write
Writing to the PCA9555 (GPIO expansion chip) failed (Replace control
board).
Limit switch head side rail right
Verify wiring to limit switch and replace limit switch.
Limit switch head side rail left
Verify wiring to limit switch and replace limit switch.
Limit switch foot side rail right
Verify wiring to limit switch and replace limit switch.
Limit switch foot side rail left
Verify wiring to limit switch and replace limit switch.
Foot Right Load cell over range
Foot Right load cell or cabling is damaged (Replace load cell).
Head Right cell over range
Head Right load cell or cabling is damaged (Replace load cell).
Foot Left Load cell over range
Foot Left load cell or cabling is damaged (Replace load cell).
Head Left Load cell over range
Head Left load cell or cabling is damaged (Replace load cell).
Motor Brake Overheat
Brake motor has ran too long (Leave motor stationary for 54 minutes).
Motor Brake Overload
Brake motor is drawing too many AMPS (remove restriction or replace
motor).
Motor Foot Overheat
Foot motor has ran too long (Leave motor stationary for 54 minutes).
Motor Foot Overload
Foot motor is drawing too many AMPS (remove restriction or replace motor).
Motor Gatch Overheat
Gatch motor has ran too long (Leave motor stationary for 54 minutes).
Motor Gatch Overload
Gatch motor is drawing too many AMPS (remove restriction or replace
motor).
Motor Head Overheat
Fowler motor has ran too long (Leave motor stationary for 54 minutes).
Motor Head Overload
Fowler motor is drawing too many AMPS (remove restriction or replace
motor).
Motor HL Foot Overheat
Foot lift motor has ran too long (Leave motor stationary for 54 minutes).
Motor HL Foot Overload
Foot lift motor is drawing too many AMPS (remove restriction or replace
motor).
Motor HL Head Overheat
Head lift motor has ran too long (Leave motor stationary for 54 minutes).
Motor HL Head Overload
Head lift motor is drawing too many AMPS (remove restriction or replace
motor).
Motor Zoom® Overheat
Zoom® motor has ran too long (Leave motor stationary for 54 minutes).
Motor Zoom® Overload
Zoom® motor is drawing too many AMPS (remove restriction or replace
motor).
No Error
Angle sensor failure or calibration issue.
One Motor Drive Short
Short on drive motor detected.
Safe WO Cmd from A1
Network A1 has a wiring issue (Check cabling).
Safe WO Cmd from A2
Network A2 has a wiring issue (Check cabling).
Safe WO Cmd from B1
Network B1 has a wiring issue (Check cabling).
Safe WO Cmd from B2
Network B2 has a wiring issue (Check cabling).
Scale ADC Error
Scale chip not calibrated. DC Control Board needs to be replaced.