Troubleshooting
Figure 50
Figure 49
Figure 51
Figure 52
SCALE TROUBLESHOOTING
When the Scale system is unable to correctly weigh the patient weight due to a problem with the electronics, Figure 49
appears. It also appears when there is a problem with the Trendelenberg angle sensor. Because of this, the value for
the weight and the angle cannot be displayed.
When the weight exceeds 550 lbs., Figure 50 appears. If the weight is less than 2 lbs., the screen displays “0 lb.”
When the Trendelenburg or Reverse Trendelenburg angle is above 12
0
or below -12
0
, Figure 51 appears.
When the scale system is unstable due to a problem with scale electronics or excessive patient movement, the weight
reading is orange (Figure 52).